Home » Kaye ValProbe Real-Time Wireless Data Logger
Business

Kaye ValProbe Real-Time Wireless Data Logger

Wireless thermal validation for standard processes. Kaye ValProbe RT is the successor of the widely recognized Kaye ValProbe, the accepted standard in wireless thermal validation systems for over 18 years.

Kaye ValProbe® RT (Real-Time) is a wireless thermal validation system designed around the specific thermal validation and regulatory needs of the Pharmaceutical and Biotech industries.

As the reliability, repeatability and accuracy of real-time wireless technology has continued to evolve, Kaye has utilized the latest advances in RF and battery technology to design the most advanced wireless thermal validation system on the market.

The ValProbe RT wireless thermal validation system is comprised of:

  • A wide range of wireless loggers for accurate, in-situ measurements of critical process temperature , pressure, and humidity measurements. (-80C to 140C).
  • A Base Station which provides reliable real-time wireless communications and collection of process data from the loggers under the harshest of conditions. Utilization of cutting-edge technologies such as antenna diversity ensure unparalleled signal strength and reliable communications and collection of real-time data.
  • A rugged, hardened, Validation Console utilizing Windows 10 LTSC and loaded with ValProbe RT software and a coreload dedicated to Validation only tasks. This provides added security and reliability while also minimizing the software validation efforts. The ValProbe RT software provides an intuitive, easy to use operator interface to perform all your validation tasks; Setups, Qualifications, Logger Verifications, Reporting and Data Storage/ Backup. The software is designed to meet all current regulatory requirements and/or recommendations for 21 CFR Part 11, Data Integrity, and Cyber Security.

The ValProbe RT family of real-time data loggers provide accurate, convenient, and reliable process measurement for a wide range of pharmaceutical, and medical device applications.

The ValProbe RT system combines high accuracy measurements, automated sensor verification, an intuitive metro style user interface, and extensive reporting to simplify the complete validation process.

Take the next step and see how the Kaye ValProbe® RT wireless thermal validation system greatly simplifies monitoring and validation of severe and hard-to-reach environments while also providing validation efficiency and time savings.

Features

  • Life data under harsh conditions via RF during validation study
  • RF-Range up to 150 meters
  • Temperature Range -85 to 400°C
  • Up to 100,000 samples per sensor
  • Scan rate down to 1 second
  • ValProbe system capacity 25 loggers / 50 sensors
  • Customer interchangeable batteries
  • Accuracy up to 0.1°C
  • Unmatched battery life

Applications

  • Steam Sterilizer
  • Dry Heat Sterilizers
  • Steam in Place (SIP)
  • Water Cascade/Fall Sterilizer
  • Incubators
  • Stability Chambers
  • Freezers
  • Freeze Dryer/Lyophilization
  • Vessels

Specifications

Kaye ValProbe RT Generic Specifications
Base Station Dimensions  7.6 in x 5.2 in x 2.2 in
190mm x 130 mm x 55 mm
Logger Dimensions Hight: 1,9” / Diameter 1,4”
48mm / 36mm
Logger Material Stainless Steel 316L and Peek
Battery Field replaceable – 3.6V Lithium
Sampling Rate Starting at 1 sec
Data Storage 100.000 Samples
retained in non-volatile memory
Real-Time Clock Accuracy < 15sec/day
Kaye ValProbe RT Generic Specifications
Calibration NVLAP/DAkkS Calibration
Verification Automated User
Verification capability
Sensing Element Precision Platinum RTD
Environmental Temperature -85°C to 140°C
Environmental Pressure 0-5bar absolute
Environmental Humidity 0-100% condensing
Regulatory Compliance UL and CE
Temperature Loggers Pressure & Temp. Logger
Specifications Rigid Bendable Flexible Surface Pressure
Sensor Type Single Sensor Single and Dual Sensor Single, Dual and 5-Channel Ultra Flat Surface Sensor Single P/T Sensor
Sensor Length 1.5, 3, 6, 9″ inches
38, 76, 152, 229 mm
12, 24, 36″ inches
305, 610, 915 mm
40″ inches
1000 mm
Tip Diameter 0.118″; 3mm 0.095″; 2.4mm 0.095″; 2.4mm 32mm diameter 1/4 NPT Connection fitting
Measurement Range -85°C to 140°C -85°C to 400°C -85°C to 140°C -85°C to 140°C 0°C to 140°C / 0-5bar abs.
Accuracy -85°C to 140°C,     ±0.1°C -85°C to 140°C,    ±0.1°C
140°C to 400°C    ±0.25°C
-85°C to 140°C,     ±0.1°C -85°C to 140°C,     ±0.1°C 0°C to 120°C         ±25mb
120°C to 135°C     ±10mb
135°C to 140°C     ±25mb
0°C to 140°C,       ±0.1°C